ECGbook.com

Making Medical Education Free for All

ECGbook.com

Making Medical Education Free for All

Home /

Voltage Calibration of the ECG

Voltage calibration of the ECG

ECG Device and Millivolts

ECG electrode and calibration

Millivolts and the ECG Curve

  • During the cardiac cycle, the heart generates electric vectors
  • ECG electrodes placed on the patient's body detect the electric voltage of the vectors in millivolts
  • ECG device
    • Converts millivolts into an ECG curve
    • Has 10 electrodes placed on the body, which detect electric vectors
    • Displays the vectors on ECG paper in 12 ECG leads

Electrical Vectors and ECG Calibration

Electrical voltage vector and ECG calibration

12-Lead ECG and Calibration

ECG sinus rhythm with standard calibration 10mm/mV

Calibration Mark and ECG Paper

  • ECG paper almost always has a calibration mark at the beginning
  • Calibration mark
    • Is a vertical rectangle (5x10mm) at the start of the paper
    • provides information about the speed of the ECG paper and the height of the 1mV amplitude

Calibration Mark

ECG standard calibration 10mm x 5mm, 10mm = 1mV

ECG Amplitude

Calibration 10mm and 14mm ECG deflection
  • ECG curve amplitude (height, depth) depends on:
  • ECG curve amplitude
  • If all QRS complexes on the ECG are:
    • very high or low
    • consider calibration

  • Amplitude in mm corresponds to electrical voltage
    • 14mm = 1.4mV
    • 7mm = 0.7mV
      (with standard calibration 1mV = 10mm)

Calibration 10mm and 7mm ECG deflection

ECG Amplitude and Calibration


ECG calibration 1mV = 10mm

Calibration 10mm



ECG calibration 1mV = 5mm

Calibration 5mm



ECG calibration 1mV = 20mm

Calibration 20mm


ECG Paper Speed and Calibration

  • The standard ECG paper speed is 25mm/s
    • At this speed, the calibration mark has a width of 5mm
  • Sometimes a speed of 50mm/s is used (for diagnosing supraventricular tachycardia)
    • ECG paper moves twice as fast, so the ECG curve and calibration mark are twice as wide
    • The calibration mark has a width of 10mm

  • The following two ECG papers are from the same patient

Calibration width 5mm and ECG paper speed 25mm/s

Paper speed 25mm/s

  • Calibration mark width 5mm


Calibration width 10mm and ECG paper speed 50mm/s

Paper speed 50mm/s

  • ECG curve is twice as wide
  • Calibration mark width 10mm


Irregular ECG with no P wave, and calibration 10mm

Standard Calibration 10mm



Calibration 20mm ruling out diagnosis atrial fibrillation

Calibration 20mm

  • This is the same patient as in the previous ECG
    • We set the calibration to 20mm (calibration mark on the right)
    • ECG amplitude is thus twice as large
  • P waves (red arrows) are revealed before the QRS complexes


ECG left ventricular hypertrophy with standard calibration 10mm

Calibration 10mm



Dual ECG calibration. 10mm chest leads calibration, 5mm limb leads calibration

Dual Calibration 10/5mm

  • This is the same patient as in the previous ECG
  • The ECG has dual calibration (calibration mark has 2 amplitudes)
  • QRS complexes in precordial leads


ECG calibration 10mm. WPW syndrome with difficult recognize small delta wave

Calibration 10mm



ECG with clearly recognize delta wave after change caligration from 10mm to 20mm

Calibration 20mm

  • This is the same patient as in the previous ECG
    • With 20mm calibration, the ECG amplitude is 2x larger
    • The delta wave is revealed (red arrows)
  • The patient has WPW syndrome



Sources

  • ECG from Basics to Essentials Step by Step
  • litfl.com
  • ecgwaves.com
  • metealpaslan.com
  • medmastery.com
  • uptodate.com
  • ecgpedia.org
  • wikipedia.org
  • Strong Medicine
  • Understanding Pacemakers





šípka späť

Voltage Calibration of the ECG

Voltage calibration of the ECG

ECG Device and Millivolts

ECG electrode and calibration

Millivolts and the ECG Curve

  • During the cardiac cycle, the heart generates electric vectors
  • ECG electrodes placed on the patient's body detect the electric voltage of the vectors in millivolts
  • ECG device
    • Converts millivolts into an ECG curve
    • Has 10 electrodes placed on the body, which detect electric vectors
    • Displays the vectors on ECG paper in 12 ECG leads

Electrical Vectors and ECG Calibration

  • During ventricular depolarization
    • 3 electrical vectors are generated in the ventricles
      • septal, main, and terminal vectors
    • which we see on the ECG as the QRS complex

  • Each electrical vector (not just ventricular)
    • Creates an electric voltage in millivolts
  • The larger the myocardium is electrically activated
    • The larger the vector
    • The greater the voltage (millivolts)

  • Standard ECG calibration is
    • 1mV = 10mm (2 squares on top of each other)
Electrical voltage vector and ECG calibration

12-Lead ECG and Calibration

ECG sinus rhythm with standard calibration 10mm/mV

Calibration Mark and ECG Paper

  • ECG paper almost always has a calibration mark at the beginning
  • Calibration mark
    • Is a vertical rectangle (5x10mm) at the start of the paper
    • provides information about the speed of the ECG paper and the height of the 1mV amplitude

Calibration Mark

  • Indicates the height of the 1mV amplitude
    • Standardly, 1mV = 10mm

  • The width indicates the speed of the ECG paper
    • The standard paper speed is 25mm/s
    • The mark has a width of 5mm

ECG standard calibration 10mm x 5mm, 10mm = 1mV

ECG Amplitude

  • ECG curve amplitude (height, depth) depends on:
  • ECG curve amplitude
  • If all QRS complexes on the ECG are:
    • very high or low
    • consider calibration

  • Amplitude in mm corresponds to electrical voltage
    • 14mm = 1.4mV
    • 7mm = 0.7mV
      (with standard calibration 1mV = 10mm)

Calibration 10mm and 14mm ECG deflection



Calibration 10mm and 7mm ECG deflection

ECG Amplitude and Calibration


ECG calibration 1mV = 10mm

Calibration 10mm

  • 1mV = 10mm
ECG calibration 1mV = 5mm

Calibration 5mm

  • 1mV = 5mm
ECG calibration 1mV = 20mm

Calibration 20mm

  • 1mV = 20mm

ECG Paper Speed and Calibration

  • The standard ECG paper speed is 25mm/s
    • At this speed, the calibration mark has a width of 5mm
  • Sometimes a speed of 50mm/s is used (for diagnosing supraventricular tachycardia)
    • ECG paper moves twice as fast, so the ECG curve and calibration mark are twice as wide
    • The calibration mark has a width of 10mm

  • The following two ECG papers are from the same patient

Calibration width 5mm and ECG paper speed 25mm/s Calibration width 10mm and ECG paper speed 50mm/s

Paper speed 25mm/s

  • Calibration mark width 5mm

Paper speed 50mm/s

  • ECG curve is twice as wide
  • Calibration mark width 10mm


Irregular ECG with no P wave, and calibration 10mm

Standard Calibration 10mm



Calibration 20mm ruling out diagnosis atrial fibrillation

Calibration 20mm

  • This is the same patient as in the previous ECG
    • We set the calibration to 20mm (calibration mark on the right)
    • ECG amplitude is thus twice as large
  • P waves (red arrows) are revealed before the QRS complexes


ECG left ventricular hypertrophy with standard calibration 10mm

Calibration 10mm



Dual ECG calibration. 10mm chest leads calibration, 5mm limb leads calibration

Dual Calibration 10/5mm

  • This is the same patient as in the previous ECG
  • The ECG has dual calibration (calibration mark has 2 amplitudes)
  • QRS complexes in precordial leads


ECG calibration 10mm. WPW syndrome with difficult recognize small delta wave

Calibration 10mm



ECG with clearly recognize delta wave after change caligration from 10mm to 20mm

Calibration 20mm

  • This is the same patient as in the previous ECG
    • With 20mm calibration, the ECG amplitude is 2x larger
    • The delta wave is revealed (red arrows)
  • The patient has WPW syndrome



Sources

  • ECG from Basics to Essentials Step by Step
  • litfl.com
  • ecgwaves.com
  • metealpaslan.com
  • medmastery.com
  • uptodate.com
  • ecgpedia.org
  • wikipedia.org
  • Strong Medicine
  • Understanding Pacemakers