ECGbook.com

Making Medical Education Free for All

ECGbook.com

Making Medical Education Free for All

Home /

Pacemaker Intervals, Sensing, and Pacing

Pacemaker timing cycles, Sensing, Pacing

Sensing

  • It is the ability of the pacemaker to sense intrinsic electrical impulses of the heart (P waves, QRS complexes) through the electrode
  • Impulses from the chamber are the only input information for the pacemaker, which then
    • evaluates their presence (including frequency and regularity)
  • Based on sensing, the pacemaker triggers pacing
  • Placing a magnet on the chest over the implanted pacemaker disables sensing

  • Sensing sensitivity
    • It is the ability to detect intrinsic electrical activity (P wave, QRS)
    • Sensing issues include:
      • Undersensing
      • Oversensing

Sensing and VVI Pacemaker

Pacemaker VVI mode

Pacemaker normal sensing

Normal Sensing

  • Sensing is set to 3mV
  • The intrinsic QRS complex exceeded the sensitivity threshold
  • Sensing is positive


Pacemaker undersensing

Undersensing



Pacemaker oversensing

Oversensing

  • Sensing is set to 1.5mV
  • The intrinsic QRS complex and the P wave exceeded the sensitivity threshold
  • Sensing is doubly positive (which is an error)
    • Because the VVI pacemaker should only be sensing the ventricle (QRS)

Pacing



Heart atrial pacing electrode - lead
ECG spike, atrial pacing

Atrial Pacing

  • The pacemaker stimulates the atria
  • The artifact is before the P wave
  • AAI mode


ECG spike, ventricular pacing
Heart ventricular pacing

Ventricular Pacing



ECG dual chamber pacing
Heart dual chamber pacing

Dual Chamber Pacing

  • The pacemaker stimulates both the atria and ventricles
  • The artifact is before both the P wave and the QRS complex
  • DDD mode

Interval and Pacemaker

ECG pacemaker interval, large square, small square

Pacemaker and Time Intervals

  • The pacemaker uses its own software to decide:
    • When to activate and deactivate sensing
    • When to activate and deactivate pacing
  • The pacemaker has defined time intervals (timing cycles) for sensing and pacing
    • These intervals are used to control sensing and pacing


ECG pacemaker timing cycles, AVI, PVARP, VRP, LRI, TARP, AEI interval

DDD Pacemaker


Lower Rate Limit (LRL), Lower Rate Interval (LRI)

  • It is the slowest heart rate that the pacemaker allows
  • Sensing
    • If the natural heart rate is higher than the LRL, the pacemaker inhibits pacing
  • Pacing
    • If the natural heart rate is lower than the LRL, pacing is activated

  • The pacemaker has intervals defined in ms
    • If the LRL is defined as 1000ms (LRI - Lower Rate Interval)
    • Then the LRL frequency is 60/min. (60,000ms / 1000ms)

Lower Rate Limit (LRL), LRI (Lower Rate Interval)

VVI Pacemaker

  • The pacemaker is in VVI mode
  • The LRL is seen as pacing artifacts (Vp - Ventricular pacing)
  • The 4th QRS is natural and occurred earlier than the programmed LRL
  • The interval after the natural QRS and the following pacing is called the escape interval

Automatic Interval and Escape Interval


ECG pacemaker, automatic interval, escape interval

VVI Pacemaker


Hysteresis

  • Hysteresis means that:
    • Positive sensing prolongs the subsequent escape interval
  • The goal is to maintain the natural sinus rhythm for as long as possible, preserving AV synchronization

ECG pacemaker, hysteresis, automatic interval, escape interval, paced QRS, spontaneous QRS complex

VVI Pacemaker

  • The pacemaker is in VVI mode
  • The pacemaker has an LRL of 1000ms (60/min.)
    • The patient has pacing at 60/min.
  • Hysteresis of 1200ms (50/min.)
  • After a natural QRS, hysteresis occurs (the escape interval is 1200ms)
    • If no natural QRS occurs within 1200ms, pacing is activated again
  • Later, during the hysteresis escape interval, a sinus contraction occurs

Pacemaker Refractory Period (ARP, PVARP, VRP)


VRP (Ventricular Refractory Period)

  • It is the period after ventricular sensing or pacing
  • The pacemaker ignores sensing during VRP
    • It serves to prevent sensing of the T wave

ECG pacemaker VVI mode, VRP - Ventricular Refractory Period, LRL - Lower Rate Limit

VVI Pacemaker and VRP (Ventricular Refractory Period)

  • The pacemaker is in VVI mode
  • VRP (Ventricular Refractory Period)
    • It is the period after ventricular sensing or pacing
    • The pacemaker ignores sensing during VRP (Vs - Ventricular sensing)
    • It serves to prevent sensing of the T wave
  • The LRL is seen as pacing artifacts (Vp - Ventricular pacing)
  • The 4th QRS is intrinsic and occurred earlier than the programmed LRL

ARP (Atrial Refractory Period)


ECG pacemaker AAI mode, ARP (Atrial Refractory Period), LRL - Lower Rate Limit, escape interval

AAI Pacemaker and ARP (Atrial Refractory Period)


AV and VA Interval

  • A dual-chamber pacemaker has a cycle defined by 3 intervals:
  • LRL (Lower Rate Limit)
    • It is the interval between ventricular pacing events
  • AV interval (AV delay)
    • It is the interval between atrial and ventricular pacing
  • VA interval (Ventricle to Atrium)
    • It is the interval between ventricular and atrial pacing
    • Sometimes referred to as the atrial escape interval


ECG pacemaker, AV delay, VA interval (Ventricle to Atrium), LRL inteval

DDD Pacemaker (ApVp)



ECG pacemaker, Lower Rate Limit, AV delay, Ventricle to Atrium

DDD Pacemaker (AsVs)

  • A dual-chamber pacemaker operates in DDD mode (AsVs)
    • Intrinsic heart activity (P and QRS) inhibits pacing
    • Pacing artifacts are not visible on the ECG, indicating complete inhibition of the pacemaker
  • LRL (Lower Rate Limit)
    • It is the interval between ventricular pacing events
    • Pacing is inhibited as the intrinsic QRS occurs before the LRL interval elapses
  • AV interval (AV delay)
    • It is the interval between atrial and ventricular pacing
    • Pacing is inhibited as the intrinsic QRS occurs before the AV interval elapses
  • VA interval (Ventricle to Atrium)
    • It is the interval between ventricular and atrial pacing
    • Pacing is inhibited as the intrinsic P wave occurs before the VA interval elapses


ECG pacemaker DDD mode ApVs, LRL, AV interval, VA interval

DDD Pacemaker (ApVs)



ECG pacemaker DDD AsVp, LRL, AV interval - delay, VA interval

DDD Pacemaker (AsVp)

  • A dual-chamber pacemaker operates in DDD mode (AsVp)
    • An intrinsic P wave is followed by a paced QRS complex
  • LRL (Lower Rate Limit)
    • It is the interval between ventricular pacing events
    • Pacing is inhibited as the paced QRS occurs before the LRL interval elapses
  • AV interval (AV delay)
    • It is the interval between atrial and ventricular pacing
    • Ventricular pacing is activated as the intrinsic QRS does not occur before the AV interval elapses
  • VA interval (Ventricle to Atrium)
    • It is the interval between ventricular and atrial pacing
    • Atrial pacing is inhibited as the intrinsic P wave occurs before the VA interval elapses

Cross Talk


ECG pacemaker cross talk, atrial pacing, ventricular sensing

Cross Talk


Blanking Period

  • It is a protection mechanism of the pacemaker against cross talk
  • After pacing or sensing, a blanking period (blind period) always follows
    • During this time, the pacemaker is "blind" (Ignores any electrical activity)

A-blank and V-blank Periods

ECG pacemaker blanking period, A blank, V blank

Upper Rate Limit (URL)

  • It is the maximum frequency at which a pacemaker can transmit P waves to the ventricles (P wave tracking)
    • Sometimes referred to as the Maximum Tracking Rate (MTR)
  • URL protects the ventricles from tachycardia
    • For example, during atrial fibrillation, impulses from the atria could be transmitted to the ventricles at a rate of 350-600/min.

ECG pacemaker URL (upper rate limit), maximum tracking rate (MTR)

DDD Pacemaker (ApVp)

  • A dual-chamber pacemaker has a DDD mode (ApVp)
  • URL (Upper Rate Limit)
    • It is the maximum frequency at which the pacemaker can transmit P waves to the ventricles (P wave tracking)
    • If the P waves have a higher frequency than the URL, the upper rate behavior is activated

Maximum Sensor Rate (MSR)


ECG pacemaker VVIR pacing mode, Maximum sensor rate (MSR)

Maximum Sensor Rate (MSR)


Post Ventricular Atrial Refractory Period (PVARP)

  • PVARP (Post Ventricular Atrial Refractory Period)
    • It is a period after ventricular pacing (or sensing) during which an atrial signal cannot trigger AV conduction
    • It serves to prevent sensing of the retrograde P wave or an atrial extrasystole

ECG pacemaker DDD mode, PVARP (Post Ventricular Atrial Refractory Period), VRP (Ventricular Refractory Period)

DDD pacemaker (ApVp)

  • PVARP (Post Ventricular Atrial Refractory Period)
    • It is a period after ventricular pacing (or sensing) during which an atrial signal cannot trigger AV conduction
    • It serves to prevent sensing of the retrograde P wave or an atrial extrasystole
  • VRP (Ventricular Refractory Period)
    • It is a period after ventricular sensing (or pacing)
    • The pacemaker ignores sensing during VRP
    • It serves to prevent sensing of the T wave


ECG pacemaker timing cycles, AV delay, V-Blanking, A-Blanking, PVARP (Post ventricular atrial refractory period), VA interval, VRP (Ventricular refractory period), URL (Upper rate limit), LRL (Lower rate limit)

DDD pacemaker (ApVp)




Sources

  • ECG from Basics to Essentials Step by Step
  • litfl.com
  • ecgwaves.com
  • metealpaslan.com
  • medmastery.com
  • uptodate.com
  • ecgpedia.org
  • wikipedia.org
  • Strong Medicine
  • Understanding Pacemakers





šípka späť

Pacemaker Intervals, Sensing, and Pacing

Pacemaker timing cycles, Sensing, Pacing

Sensing

  • It is the ability of the pacemaker to sense intrinsic electrical impulses of the heart (P waves, QRS complexes) through the electrode
  • Impulses from the chamber are the only input information for the pacemaker, which then
    • evaluates their presence (including frequency and regularity)
  • Based on sensing, the pacemaker triggers pacing
  • Placing a magnet on the chest over the implanted pacemaker disables sensing

  • Sensing sensitivity
    • It is the ability to detect intrinsic electrical activity (P wave, QRS)
    • Sensing issues include:
      • Undersensing
      • Oversensing

Sensing and VVI Pacemaker

  • We will demonstrate sensitivity using the example of a VVI pacemaker
    • V - ventricular pacing
    • V - ventricular sensing
    • I - positive sensing inhibits pacing
    • The VVI pacemaker senses intrinsic QRS through the electrode in the ventricle

Pacemaker VVI mode


Pacemaker normal sensing

Normal Sensing

  • Sensing is set to 3mV
  • The intrinsic QRS complex exceeded the sensitivity threshold
  • Sensing is positive


Pacemaker undersensing

Undersensing

  • Sensing is set to 5mV
  • The intrinsic QRS complex did not exceed the sensitivity threshold
  • Sensing is negative (which is an error)


Pacemaker oversensing

Oversensing

  • Sensing is set to 1.5mV
  • The intrinsic QRS complex and the P wave exceeded the sensitivity threshold
  • Sensing is doubly positive (which is an error)
    • Because the VVI pacemaker should only be sensing the ventricle (QRS)

Pacing



Heart atrial pacing electrode - lead ECG spike, atrial pacing

Atrial Pacing

  • The pacemaker stimulates the atria
  • The artifact is before the P wave
  • AAI mode


Heart ventricular pacing ECG spike, ventricular pacing

Ventricular Pacing

  • The pacemaker stimulates the ventricles
  • The artifact is before the QRS complex
  • VVI mode


Heart dual chamber pacing ECG dual chamber pacing

Dual Chamber Pacing

  • The pacemaker stimulates both the atria and ventricles
  • The artifact is before both the P wave and the QRS complex
  • DDD mode

Interval and Pacemaker

  • 1 min = 60 s = 60,000 ms

  • Frequency is the number of systoles/minute
    • Systole is equivalent to the QRS complex on an ECG
    • Commonly used when evaluating an ECG

  • Interval is measured in ms (milliseconds)
    • RR interval is the number of ms between RR peaks
    • Used when evaluating a pacemaker

  • Frequency = 60,000 / interval (ms)

ECG pacemaker interval, large square, small square

Pacemaker and Time Intervals

  • The pacemaker uses its own software to decide:
    • When to activate and deactivate sensing
    • When to activate and deactivate pacing
  • The pacemaker has defined time intervals (timing cycles) for sensing and pacing
    • These intervals are used to control sensing and pacing


ECG pacemaker timing cycles, AVI, PVARP, VRP, LRI, TARP, AEI interval

DDD Pacemaker


Lower Rate Limit (LRL), Lower Rate Interval (LRI)

  • It is the slowest heart rate that the pacemaker allows
  • Sensing
    • If the natural heart rate is higher than the LRL, the pacemaker inhibits pacing
  • Pacing
    • If the natural heart rate is lower than the LRL, pacing is activated

  • The pacemaker has intervals defined in ms
    • If the LRL is defined as 1000ms (LRI - Lower Rate Interval)
    • Then the LRL frequency is 60/min. (60,000ms / 1000ms)

Lower Rate Limit (LRL), LRI (Lower Rate Interval)

VVI Pacemaker

  • The pacemaker is in VVI mode
  • The LRL is seen as pacing artifacts (Vp - Ventricular pacing)
  • The 4th QRS is natural and occurred earlier than the programmed LRL
  • The interval after the natural QRS and the following pacing is called the escape interval

Automatic Interval and Escape Interval


ECG pacemaker, automatic interval, escape interval

VVI Pacemaker


Hysteresis

  • Hysteresis means that:
    • Positive sensing prolongs the subsequent escape interval
  • The goal is to maintain the natural sinus rhythm for as long as possible, preserving AV synchronization

ECG pacemaker, hysteresis, automatic interval, escape interval, paced QRS, spontaneous QRS complex

VVI Pacemaker

  • The pacemaker is in VVI mode
  • The pacemaker has an LRL of 1000ms (60/min.)
    • The patient has pacing at 60/min.
  • Hysteresis of 1200ms (50/min.)
  • After a natural QRS, hysteresis occurs (the escape interval is 1200ms)
    • If no natural QRS occurs within 1200ms, pacing is activated again
  • Later, during the hysteresis escape interval, a sinus contraction occurs

Pacemaker Refractory Period (ARP, PVARP, VRP)


VRP (Ventricular Refractory Period)

  • It is the period after ventricular sensing or pacing
  • The pacemaker ignores sensing during VRP
    • It serves to prevent sensing of the T wave

ECG pacemaker VVI mode, VRP - Ventricular Refractory Period, LRL - Lower Rate Limit

VVI Pacemaker and VRP (Ventricular Refractory Period)

  • The pacemaker is in VVI mode
  • VRP (Ventricular Refractory Period)
    • It is the period after ventricular sensing or pacing
    • The pacemaker ignores sensing during VRP (Vs - Ventricular sensing)
    • It serves to prevent sensing of the T wave
  • The LRL is seen as pacing artifacts (Vp - Ventricular pacing)
  • The 4th QRS is intrinsic and occurred earlier than the programmed LRL

ARP (Atrial Refractory Period)


ECG pacemaker AAI mode, ARP (Atrial Refractory Period), LRL - Lower Rate Limit, escape interval

AAI Pacemaker and ARP (Atrial Refractory Period)


AV and VA Interval

  • A dual-chamber pacemaker has a cycle defined by 3 intervals:
  • LRL (Lower Rate Limit)
    • It is the interval between ventricular pacing events
  • AV interval (AV delay)
    • It is the interval between atrial and ventricular pacing
  • VA interval (Ventricle to Atrium)
    • It is the interval between ventricular and atrial pacing
    • Sometimes referred to as the atrial escape interval


ECG pacemaker, AV delay, VA interval (Ventricle to Atrium), LRL inteval

DDD Pacemaker (ApVp)



ECG pacemaker, Lower Rate Limit, AV delay, Ventricle to Atrium

DDD Pacemaker (AsVs)

  • A dual-chamber pacemaker operates in DDD mode (AsVs)
    • Intrinsic heart activity (P and QRS) inhibits pacing
    • Pacing artifacts are not visible on the ECG, indicating complete inhibition of the pacemaker
  • LRL (Lower Rate Limit)
    • It is the interval between ventricular pacing events
    • Pacing is inhibited as the intrinsic QRS occurs before the LRL interval elapses
  • AV interval (AV delay)
    • It is the interval between atrial and ventricular pacing
    • Pacing is inhibited as the intrinsic QRS occurs before the AV interval elapses
  • VA interval (Ventricle to Atrium)
    • It is the interval between ventricular and atrial pacing
    • Pacing is inhibited as the intrinsic P wave occurs before the VA interval elapses


ECG pacemaker DDD mode ApVs, LRL, AV interval, VA interval

DDD Pacemaker (ApVs)



ECG pacemaker DDD AsVp, LRL, AV interval - delay, VA interval

DDD Pacemaker (AsVp)

  • A dual-chamber pacemaker operates in DDD mode (AsVp)
    • An intrinsic P wave is followed by a paced QRS complex
  • LRL (Lower Rate Limit)
    • It is the interval between ventricular pacing events
    • Pacing is inhibited as the paced QRS occurs before the LRL interval elapses
  • AV interval (AV delay)
    • It is the interval between atrial and ventricular pacing
    • Ventricular pacing is activated as the intrinsic QRS does not occur before the AV interval elapses
  • VA interval (Ventricle to Atrium)
    • It is the interval between ventricular and atrial pacing
    • Atrial pacing is inhibited as the intrinsic P wave occurs before the VA interval elapses

Cross Talk


ECG pacemaker cross talk, atrial pacing, ventricular sensing

Cross Talk


Blanking Period

  • It is a protection mechanism of the pacemaker against cross talk
  • After pacing or sensing, a blanking period (blind period) always follows
    • During this time, the pacemaker is "blind" (Ignores any electrical activity)

A-blank and V-blank Periods

  • Atrial Blanking Period (A-blank)
    • It is a blind period after ventricular pacing
    • The atrial electrode ignores electrical activity
  • Ventricular Blanking Period (V-blank)
    • It is a blind period after atrial pacing
    • The ventricular electrode ignores electrical activity

  • A-blank and V-blank periods serve as protection against cross talk

ECG pacemaker blanking period, A blank, V blank

Upper Rate Limit (URL)

  • It is the maximum frequency at which a pacemaker can transmit P waves to the ventricles (P wave tracking)
    • Sometimes referred to as the Maximum Tracking Rate (MTR)
  • URL protects the ventricles from tachycardia
    • For example, during atrial fibrillation, impulses from the atria could be transmitted to the ventricles at a rate of 350-600/min.

ECG pacemaker URL (upper rate limit), maximum tracking rate (MTR)

DDD Pacemaker (ApVp)

  • A dual-chamber pacemaker has a DDD mode (ApVp)
  • URL (Upper Rate Limit)
    • It is the maximum frequency at which the pacemaker can transmit P waves to the ventricles (P wave tracking)
    • If the P waves have a higher frequency than the URL, the upper rate behavior is activated

Maximum Sensor Rate (MSR)


ECG pacemaker VVIR pacing mode, Maximum sensor rate (MSR)

Maximum Sensor Rate (MSR)


Post Ventricular Atrial Refractory Period (PVARP)

  • PVARP (Post Ventricular Atrial Refractory Period)
    • It is a period after ventricular pacing (or sensing) during which an atrial signal cannot trigger AV conduction
    • It serves to prevent sensing of the retrograde P wave or an atrial extrasystole

ECG pacemaker DDD mode, PVARP (Post Ventricular Atrial Refractory Period), VRP (Ventricular Refractory Period)

DDD pacemaker (ApVp)

  • PVARP (Post Ventricular Atrial Refractory Period)
    • It is a period after ventricular pacing (or sensing) during which an atrial signal cannot trigger AV conduction
    • It serves to prevent sensing of the retrograde P wave or an atrial extrasystole
  • VRP (Ventricular Refractory Period)
    • It is a period after ventricular sensing (or pacing)
    • The pacemaker ignores sensing during VRP
    • It serves to prevent sensing of the T wave


ECG pacemaker timing cycles, AV delay, V-Blanking, A-Blanking, PVARP (Post ventricular atrial refractory period), VA interval, VRP (Ventricular refractory period), URL (Upper rate limit), LRL (Lower rate limit)

DDD pacemaker (ApVp)




Sources

  • ECG from Basics to Essentials Step by Step
  • litfl.com
  • ecgwaves.com
  • metealpaslan.com
  • medmastery.com
  • uptodate.com
  • ecgpedia.org
  • wikipedia.org
  • Strong Medicine
  • Understanding Pacemakers